FAQs for obsolete products
How do the valves in Scanivalve Modules
operate?
All ZOC16, ZOC17, and DSA modules have four way valves for each
sensor. Application of a control pressure causes a piston in each
valve to compress an o-ring, forming an air tight seal. All of
these valves are capable of, and have been tested to, a minimum of
one million operations.
All ZOC 22, 23 and 33 modules also have four way valves for each
sensor. These valves use inflatable bladders instead of pistons.
Although the method is somewhat different, the bladders are able to
form air tight seals.
What are control pressures?
Control pressures are used to switch the valves in a module to one
of the standard operating modes: Zero, Operate, Calibrate, and
Isolate. Electronic signals are not required to switch these
valves.
How do the control pressures operate the
valves?
Each module requires two or three control pressures to operate the
valves, depending on the valve logic. ZOC16, ZOC17 and all DSA
modules require three pressures. Two of the pressures control the
normal valve modes. The third pressure controls a separate purge
valve.
ZOC22, ZOC23, and ZOC33 modules also require two or three
control pressures. The standard versions of these modules require
two control pressures. The Duplex versions of these modules require
a third control pressure to operate the Duplex valve.
The different modes of operation are obtained by applying the
control pressures in a logical sequence. Truth Tables may be found
in the appropriate module hardware manual that show the logic
required for proper operation.
Do control pressures have to be applied at all
times?
All DSA series, ZOC16, and ZOC17 modules are" normally open"
valves. That is, no control pressure is required to put the module
in the Operate mode. Control pressures are required to switch the
valves to the other modes of operation. The control pressure used
for these modules must be 85 to 105 psi depending on the pressure
range of the module.
All ZOC22, ZOC23, and ZOC33 modules require a control pressure
for each mode of operation. The control pressure for these modules
must be 60 to 65 psi.
Is it possible to minimize the number of control pressures
needed?In some cases, yes. For DSA modules, a single control
pressure may be used for "Zero Calibrations" Also, CTL 1 and CTL 2
may be applied from the same source to switch the module into the
Calibration mode. ZOC22, ZOC23, and ZOC33 modules must have all
control pressures available.
What quality of air should be used for control
pressures?
In order to prevent damage to the valves and sensors, the control
pressures for all modules must be "Instrument Grade". For more
information, please refer to ISA Specification ISA-S7-3.
How do I upgrade the software?
Software upgrades are available from Scanivalve Corp. A detailed
upgrade procedure may be found in the DSA3200 Series Software
Specification, which is also available for download from this
website. For more information on updating software for these
modules, contact Scanivalve Corp, Product Service Department.
I have lost my backup coefficients. What can I
do?
Scanivalve Corp keeps a record of the coefficients for each module
generated in the most recent calibration performed at Scanivalve
Corp. These coefficients can be obtained at no charge by contacting
Scanivalve Corp, Product Support Dept.
The original coefficients for any module calibrated with a
customer DSM is stored in a backup folder on the DSM hard drive. In
addition, the coefficients are shipped on a floppy disk as a
backup. If the coefficients cannot be found, copies of the last
calibration coefficients may be obtained by contacting Scanivalve
Corp, Product Support Department.
How do I obtain coefficients for my
modules?
Scanivalve Corp keeps a record of the coefficients for each module
generated in the most recent calibration performed at Scanivalve
Corp. These coefficients can be obtained at no charge by contacting
Scanivalve Corp, Product Support Dept.
How do I upgrade the software?
Software upgrades are available from Scanivalve Corp. A detailed
procedure may be found in the RAD3200 Manual, which is also
available for download from this website. For more information on
updating software for these modules, contact Scanivalve Corp,
Product Service Department.
How do I upgrade the software?
Software upgrades are available from Scanivalve Corp. A detailed
procedure may be found in the DSAENCL3000 Manual, which is also
available for download from this website. For more information on
updating software for these modules, contact Scanivalve Corp,
Product Service Department.
SPC3000 Calibrator Modules
SPC3000 Calibrator Modules are obsolete. They have been replaced
by the SPC4000 Calibrator.
How many pressures must be applied to the
calibrator?
All servo operated calibrators require a 95 psi supply to operate
the internal valves. The Servo operates from a separate servo
supply.
How accurate must the control and servo pressures
be?
The control supply may be 85 to 105 psi depending on the full
scale pressure range of the module. The servo supply must be 3 to
10 psi greater than the full scale range of the module. The servo
supply must be very well regulated.
What quality of air must be applied to the
calibrators?
In order to prevent damage to the valves and secondary standard,
the control and servo pressures must be "Instrument Grade". For
more information, please refer to ISA Specification ISA-S7-3.
My calibrator oscillates, what do I do?
The stability of a Servo operated calibrator depends upon the
servo supply input and the volume connected to the output of the
calibrator. All servo operated calibrators are "tuned" at the
Scanivalve Factory prior to shipment. Unless the customer
installation exactly matches the Scanivalve Factory air supply, the
modules may have to be re-tuned prior to use. The modules may be
re-tuned by executing a "TUNE" command.
I cannot communicate with my calibrator.
All calibrators communicate with a Scanivalve Pressure Measurement
System or Host computer using a RS-232 connection. The settings
are: 9600 BAUD, 8 data bits, 1 stop bit, no parity. All calibrators
are designed to be used in systems with multiple calibrators in a
"daisy chain" configuration. Each calibrator has a unique address
from 1 to 9. Commands to the calibrator must be preceded by the
calibrator address. All calibrators shipped from the Scanivalve
Factory have the address set to 1 unless the calibrator is part of
a larger system. The address may be changed by a user by changing
the setting of ADDR. For more information consult the Calibrator
manual or contact Scanivalve Corp, Product Support Dept.
How often should I calibrate the module?
Calibrators with Mensor Secondary Standards DSA modules should be
re-calibrated at a six month interval for the best accuracy.
Calibrators with a Digiquartz Secondary Standard should be
re-calibrated at six month intervals for the first year. Generally
after the first year the calibration interval of Digiquartz
Secondary Standards may be extended to one year.
ZOC Pressure Modules
What are the power requirements?
All ZOC modules require + 15Vdc and -15 Vdc for operation. This
voltage does not need to be instrument grade. Sensor excitation
voltage is developed on board the module.
How do I address the modules?
Addressing is a 4 to 6 bit address line depending on the number of
sensors in the module. All address line inputs are pulled up to the
+15 Vdc input.
How do I calibrate the modules?
The modules may be calibrated by applying the correct control
pressure to switch the module to the Calibrate mode. Calibration
pressures may be applied to all sensors from the CAL input. It is
recommended that a second order polynomial be used to characterize
each sensor for the best accuracy.
How often should I calibrate the modules?
ZOC pressure modules will show a considerable amount of drift
unless the module is installed in a Thermal Control Unit. For best
accuracy, a ZOC module should be re-zeroed before each data point.
Calibration should be checked prior to taking data. Scanivalve Corp
provides a utility program for customers who can perform their own
calibrations. This program is called PressCal and may be downloaded
from this website at no charge.
Mechanical Scanivalves/Fluid Switch
Wafers
Scanivalve Corp regrets that the entire Mechanical Scanivalve
product line can no longer be supported. Components for the
products in this line cannot be obtained nor manufactured. The
products in this line were designed prior to 1980 and obsoleted in
1995.
My Scanivalve will not step.
Scanivalves with solenoid drives require 24 Vdc at 4 to 6 amps to
step.
Scanivalves with stepper motors require a step or home contact
closure. These systems have opto-isolated inputs that require a
pull up voltage. If this voltage is present, the power supply may
be at fault.
My Scanivalve System steps, but it is very
rough.
The motor driver board or the power supply board may have to be
replaced.
My Scanivalve System tries to step, but it is cannot
complete the step.
The motor driver board or the motor logic board may have to be
replaced.
My Scanivalve cannot find Home.
Odd-Even or other encoders in Type J, D, or S Scanivalves may have
to be aligned. Scanivalve systems also may need encoder
re-alignments, but they could also have a failed optical diode.
How do I interface to the Scanivalve
Encoder.
Encoder circuits wiring diagrams may be downloaded from the
Product Information Section of this website.
How do I connect the wires on a Solenoid
Drive?
Solenoid Drive wiring diagrams may be downloaded from the Product
Information Section of this website.
What is the recommended Lubrication and Maintenance
schedule for Scanivalves?
This information may be downloaded from the Product Information
Section of this website.
What is Balance Pressure and how should it be
applied?
Older Type D, Type S, and all Fluid Switch Wafers(except model
W62) require balance pressure for proper operation. The balance
pressure should be equal to the average input pressure. Also, input
pressures should be equally distributed around the inputs to
prevent very high pressures in one quadrant. If the input pressures
are below ambient pressure, Balance pressure may not need to be
applied unless the pressures will be greater than 3 psi below
ambient. At that time a low vacuum pressure may need to be
applied.
Scanivalve Digital Interface Units
Scanivalve Corp regrets that the Scanivalve Digital Interface Unit
(SDIU) product line can no longer be supported. Components for the
products in this line cannot be obtained nor manufactured. The
products in this line were designed prior to 1985 and obsoleted in
1995.
The SDIU will not accept commands from a Host
computer.
The "End of Transmission" characters stored in memory do not match
the characters used by the Host Computer.
The SDIU loses the setup when powered
down.
The battery in the Non-volatile RAM chip has failed. The
chip(Dallas DS1220) must be replaced.
The display is flickering.
The power supply voltage(s) is(are) too low.