Frequently Asked Questions

Digital Thermocouple Modules
Q: I cannot communicate with the Unit on the Ethernet connection. Answer
Q: How do I change the IP address?  Answer
Q: How do I upgrade the software?  Answer
Q: What are the power requirements?  Answer
Q: How often should I calibrate the module?  Answer


I cannot communicate with the Unit on the Ethernet connection.

Verify that the Ethernet cable is the correct cable for the connection. A Straight wired@ cable must be used if the Host computer is connected to the DTS module through a hub. For direct connections from a Host to a DTS module, a Across wired@ cable must be used.

The Host computer must have an IP address specified to be in the same class as the DTS module. That is, if the DTS module has an IP address of 191.30.5.100, the Host computer must have an IP address of 191.30.x.y.

Read FAQs on Digital Thermocouple Modules
View all FAQ topics

How do I change the IP address?
The IP Address in a DTS3250 series module may be changed by a user. The IP address is stored as a variable in the operating system. This variable can be changed only by using the RS232 connection. A procedure for this may be found in the DTS3250 Manual

Read FAQs on Digital Thermocouple Modules
View all FAQ topics

How do I upgrade the software?
Software upgrades are available from Scanivalve Corp. A detailed procedure may be found in the DTS3250 Manual, which is also available for download from this website. For more information on updating software for these modules, contact Scanivalve Corp, Product Service Department.

Read FAQs on Digital Thermocouple Modules
View all FAQ topics

What are the power requirements?
The DTS3250/16Tx module with a heater requires 24 Vdc @ 2.5A during warmup. Once the module has reached temperature, the current requirement will drop to 0.5A. This same module less heater requires 24Vdc @ 0.5A maximum.

Read FAQs on Digital Thermocouple Modules
View all FAQ topics

How do I upgrade the software?
Software upgrades for these modules can only be made by changing a WSI chip. This chip should only be changed by qualified technicians. For more information on updating software for these modules, contact Scanivalve Corp, Product Service Department.

Read FAQs on Digital Thermocouple Modules
View all FAQ topics

How often should I calibrate the module?
DTS modules should be re-calibrated at a six month interval for the best accuracy.

Read FAQs on Digital Thermocouple Modules
View all FAQ topics

 