Q:
I cannot communicate with the Unit on the Ethernet
connection. Answer
Q:
How do I change the IP address?
Answer
Q:
How do I upgrade the software?
Answer
Q:
What are the power requirements?
Answer
Q:
How often should I calibrate the
module? Answer
I
cannot communicate with the Unit on the Ethernet
connection.
Verify that the Ethernet cable is the correct
cable for the connection. A Straight wired@ cable
must be used if the Host computer is connected
to the DTS module through a hub. For direct connections
from a Host to a DTS module, a Across wired@ cable
must be used.
The Host computer must have an IP address specified
to be in the same class as the DTS module. That
is, if the DTS module has an IP address of 191.30.5.100,
the Host computer must have an IP address of 191.30.x.y.
Read
FAQs on Digital Thermocouple Modules
View
all FAQ topics
How
do I change the IP address?
The IP Address in a DTS3250 series module may
be changed by a user. The IP address is stored
as a variable in the operating system. This variable
can be changed only by using the RS232 connection.
A procedure for this may be found in the DTS3250
Manual
Read
FAQs on Digital Thermocouple Modules
View
all FAQ topics
How
do I upgrade the software?
Software upgrades are available from Scanivalve
Corp. A detailed procedure may be found in the
DTS3250 Manual, which is also available for download
from this website. For more information on updating
software for these modules, contact Scanivalve
Corp, Product Service Department.
Read
FAQs on Digital Thermocouple Modules
View
all FAQ topics
What
are the power requirements?
The DTS3250/16Tx module with a heater requires
24 Vdc @ 2.5A during warmup. Once the module has
reached temperature, the current requirement will
drop to 0.5A. This same module less heater requires
24Vdc @ 0.5A maximum.
Read
FAQs on Digital Thermocouple Modules
View
all FAQ topics
How
do I upgrade the software?
Software upgrades for these modules can only be
made by changing a WSI chip. This chip should
only be changed by qualified technicians. For
more information on updating software for these
modules, contact Scanivalve Corp, Product Service
Department.
Read
FAQs on Digital Thermocouple Modules
View
all FAQ topics
How
often should I calibrate the module?
DTS modules should be re-calibrated at a six month
interval for the best accuracy.
Read
FAQs on Digital Thermocouple Modules
View