What
are the power requirements?
The RADBASE requires : + 15 Vdc @ 41 mA, -15 Vdc
@ 4mA, and + 5 Vdc @ 610 mA. Each A/D Module requires:
+ 15 Vdc @ 105 mA and - 15 Vdc @ 5.5 mA. A typical
ZOC Module connected to an A/D will require:v+
15 Vdc @ 120 mA and - 15 Vdc @ 16 mA..
Read
FAQs on RAD3200/AD3200 Modules
View
all FAQ topics
The
program has stopped responding but the RAD appears
to still be scanning?
The USB connection has been compromised. The cable
is too long or not shielded properly. If a USB
extender is being used, the power may be too low.
Read
FAQs on RAD3200/AD3200 Modules
View
all FAQ topics
The
Host computer cannot connect to itself without
a Hub?
Windows XP will shut down the Ethernet software
if nothing is connected to save power. Disable
this shutdown feature in the Power Management
section of the LAN setup.
Read
FAQs on RAD3200/AD3200 Modules
View
all FAQ topics
How
do I upgrade the software?
Software upgrades are available from Scanivalve
Corp. A detailed procedure may be found in the
DSM3200 Manual, which is also available for download
from this website. For more information on updating
software for these modules, contact Scanivalve
Corp, Product Service Department.