DSA3200 Digital Sensor Array Modules
Q: I cannot communicate with the Unit on the Ethernet connection.  Answer
Q: How do I change the IP address?  Answer
Q: How do I upgrade the software?  Answer
Q: I have lost my backup coefficients. What can I do?  Answer
Q: What are the power requirements?  Answer
Q: How often should I calibrate the module?  Answer


I cannot communicate with the Unit on the Ethernet connection.

Verify that the Ethernet cable is the correct cable for the connection. A straight wired cable must be used if the Host computer is connected to the DSA module through a hub. For direct connections from a Host to a DSA module, a Across wired@ cable must be used.

The Host computer must have an IP address specified to be in the same class as the DSA module. That is, if the DSA module has an IP address of 191.30.5.100, the Host computer must have an IP address of 191.30.80.sn.

Read FAQs on DSA3200 Digital Sensor Array Modules
View all FAQ topics

How do I change the IP address?
The IP Address in a DSA3200 series module may be changed by a user. The IP address is stored as a variable in the operating system. This variable can be changed only by using the RS232 connection. A procedure for this may be found in the DSA3200 Series Software Specification.

Read FAQs on DSA3200 Digital Sensor Array Modules
View all FAQ topics

How do I upgrade the software?
Software upgrades are available from Scanivalve Corp. A detailed upgrade procedure may be found in the DSA3200 Series Software Specification, which is also available for download from this website. For more information on updating software for these modules, contact Scanivalve Corp, Product Service Department.

Read FAQs on DSA3200 Digital Sensor Array Modules
View all FAQ topics

I have lost my backup coefficients. What can I do?
Scanivalve Corp keeps a record of the coefficients for each module generated in the most recent calibration performed at Scanivalve Corp. These coefficients can be obtained at no charge by contacting Scanivalve Corp, Product Support Dept.

Read FAQs on DSA3200 Digital Sensor Array Modules
View all FAQ topics

What are the power requirements?
DSA327 and DSA3207 modules require 28 +/- 4 Vdc at 10 Watts
DSA3016 modules are powered by the DSA3016 enclosure
DSA3218 modules require 28 +/-4 Vdc at 35 Watts

Read FAQs on DSA3200 Digital Sensor Array Modules
View all FAQ topics

How often should I calibrate the module?
DSA modules should be re-calibrated at a six month interval for the best accuracy.

Read FAQs on DSA3200 Digital Sensor Array Modules
View all FAQ topics

 