Frequently
Asked Questions
DSA3000
Digital Sensor Array Modules
Q:
I cannot communicate with the Unit on the Ethernet
connection. Answer
Q:
How do I change the IP address?
Answer
Q:
I have lost my backup coefficients.
What can I do? Answer
Q:
What are the power requirements?
Answer
Q:
How do I upgrade the software?
Answer
Q:
How often should I calibrate the
module? Answer
I
cannot communicate with the Unit on the Ethernet
connection.
Verify that the Ethernet cable is the correct
cable for the connection. A straight wired cable
must be used if the Host computer is connected
to the DSA module through a hub. For direct connections
from a Host to a DSA module, a cross wired cable
must be used.
The Host computer must have an IP address specified
to be in the same class as the DSA module. That
is, if the DSA module has an IP address of 191.30.5.100,
the Host computer must have an IP address of 191.30.5.y.
Read
FAQs on DSA3000 Digital Sensor Array Modules
View
all FAQ topics
How
do I change the IP address?
The IP Address in a DSA3000 series module may
be changed by a user. The IP address is stored
as a configuration variable. This variable is
read only on power up. When an IP address is changed,
it will not become effective until module power
has been cycled.
Read
FAQs on DSA3000 Digital Sensor Array Modules
View
all FAQ topics
I
have lost my backup coefficients. What can I do?
Scanivalve Corp keeps a record of the coefficients
for each module generated in the most recent calibration
performed at Scanivalve Corp. These coefficients
can be obtained at no charge by contacting Scanivalve
Corp, Product Support Dept.
Read
FAQs on DSA3000 Digital Sensor Array Modules
View
all FAQ topics
What
are the power requirements?
DSA3017 and DSA3007 modules require 28 +/- 4 Vdc
at 8 Watts
DSA3016 modules are powered by the DSA3016 enclosure
DSA3018 modules require 28 +/-4 Vdc at 35 Watts
Read
FAQs on DSA3000 Digital Sensor Array Modules
View
all FAQ topics
How
do I upgrade the software?
Software upgrades for these modules can only be
made by changing a WSI chip. This chip should
only be changed by qualified technicians. For
more information on updating software for these
modules, contact Scanivalve Corp, Product Service
Department.
Read
FAQs on DSA3000 Digital Sensor Array Modules
View
all FAQ topics
How
often should I calibrate the module?
DSA modules should be re-calibrated at a six month
interval for the best accuracy.
Read
FAQs on DSA3000 Digital Sensor Array Modules
View
all FAQ topics
|